Here is a chart to show you our Electromagnetic flow meter flow range & pressure rating for your reference.
Normally our turn down ratio is 10:1; 30:1 is also available upon your requirement. Please specify it clearly when you need it and it is our SE12 type high accuracy electromagnetic flow meter.
Size(DN) | Pressure(Mpa) | Min flow range velocity(0-0.5)m/s | Max Flow range velocity(0-10)m/s |
10 | 4.0 | (0-2.25)L/min | (0-45)L/min |
15 | 4.0 | (0-5)L/min | (0-100)L/min |
20 | 4.0 | (0-7.5)L/min | (0-150)L/min |
25 | 4.0 | (0-10L)/min | (0-200)L/min |
32 | 4.0 | (0-20L)/min | (0-400)L/min |
40 | 4.0 | (0-30L)/min | (0-600)L/min |
50 | 4.0 | (0-3)m³/h | (0-60)m³/h |
65 | 4.0 | (0-6)m³/h | (0-120)m³/h |
80 | 4.0 | (0-9)m³/h | (0-180)m³/h |
100 | 1.6 | (0-12)m³/h | (0-240)m³/h |
125 | 1.6 | (0-21)m³/h | (0-420)m³/h |
150 | 1.6 | (0-30)m³/h | (0-600)m³/h |
200 | 1.6 | (0-54)m³/h | (0-1080)m³/h |
250 | 1.6 | (0-90)m³/h | (0-1800)m³/h |
300 | 1.0 | (0-120)m³/h | (0-2400)m³/h |
350 | 1.0 | (0-165)m³/h | (0-3300)m³/h |
400 | 1.0 | (0-225)m³/h | (0-4500)m³/h |
500 | 1.0 | (0-330)m³/h | (0-6600)m³/h |
600 | 1.0 | (0-480)m³/h | (0-9600)m³/h |
700 | 1.0 | (0-660)m³/h | (0-13200)m³/h |
800 | 1.0 | (0-900)m³/h | (0-18000)m³/h |
900 | 1.0 | (0-1200)m³/h | (0-24000)m³/h |
1000 | 1.0 | (0-1350)m³/h | (0-27000)m³/h |
1200 | 0.6 | (0-2100)m³/h | (0-42000)m³/h |
1400 | 0.6 | (0-2700)m³/h | (0-54000)m³/h |
1600 | 0.6 | (0-3600)m³/h | (0-72000)m³/h |
1800 | 0.6 | (0-4500)m³/h | (0-90000)m³/h |
2000 | 0.6 | (0-5700)m³/h | (0-114000)m³/h |